Fox Flow Meter Model FT2

Fox Flow Meter Model FT2

$0.00
Accurate Mass Measurement of Natural Gas Flow, Air flow and Other Gases

Fox Flowmeter featuredA Thermal Mass Flow Meter is an instrument used to measure linear, nonlinear, mass or volumetric flow rate of a liquid or a gas.

Fox Flow Meters use a Constant Temperature Differential (?T) technology. The sensor has two elements. The Reference RTD measures the gas temperature. The electronics heats the Heated Element above the gas temperature. It is the job of the electronics to maintain a constant ?T between the gas temperature and the Heated Element. As mass flow increases, the increased numbers of gas molecules remove more heat from the Heated Element.

The electronics senses this temperature reduction and adds additional power to the heated element in order to maintain a constant ?T. The amount of power delivered to the Heated Element is proportional to the mass flow rate. The microprocessor then linearizes this signal to deliver a linear output.

 

 

 

 

Models

The Flowmeters listed below are four common builds of Fox FT2 Flowmeters. If you are looking for a custom flowmeter that will fit the speficiations you need, download the Flowmeter Component Sheet and build your custom flowmeter.

Contact our customer service toll-free: 866-650-1937 to order your custom Fox FT2 Flowmeter. To place an order, please have your custom built Part Number using the Flowmeter Component Sheet.

Model FT2-20P-SS-ST-E1-DD-B0-G1
Fox Model FT2 Inline Flowmeter 2" 316 SS Body 24 VDC

Model FT2-20P-SS-ST-E2-DD-B0-G1
Fox Model FT2 Inline Flowmeter 2" 316 SS Body 85-250 VAC

Model FT2-06I-SS-ST-E1-DD-B0-G1
Fox Model FT2 Insertion Flowmeter 6" Probe 316 SS Body 24 VDC

Model FT2-06I-SS-ST-E2-DD-B0-G1
Fox Model FT2 Insertion Flowmeter 6" Probe 316 SS Body 85-250 VAC

Custom Order

Need to Build a Custom Flowmeter?

The Flowmeters listed below are four common builds of Fox FT2 Flowmeters. If you are looking for a custom flowmeter that will fit the speficiations you need, download the Flowmeter Component Sheet and build your custom flowmeter.

Call 866-650-1937 to order your custom Fox FT2 Flowmeter. To place an order, please have your custom built Part Number using the Flowmeter Component Sheet.

Part Number Ex: FT2-20P-SS-ST-E1-DD-B0-G1 (2" Inline Flow Meter 316 SS Body, 24 VDC )

Specifications

Download Fox Flowmeter Technical Guide

Performance Specs

Flow Accuracy:

± 1% of reading ± 0.2 % of full scale*
* Point velocity for insertion flowmeters. Fox recommends a minimum of 15 diameters of straight pipe upstream of the flowmeter and 10 diameters downstream for insertion flowmeters. Fox recommends a minimum of 8 diameters of straight pipe upstream of the flowmeter and 4 diameters downstream for inline flowmeters.

Flow Repeatability:
± 0.2% of full scale

Flow Response Time:
0.9 seconds (one time constant)

Temperature Accuracy:
± 1.8° F (± 1.0° C) over -40 to 250° F (-40 to 121° C); ± 3.6° F (± 2.0° C) over 250 to 650° F (121 to 343° C). Minimum velocity 60 SFPM.a

 


 

Operating Specs

Units of Measurement:

SCFM, SCFH, NM3/hr, NM3/min, Kg/hr, Kg/min, Kg/sec, Lbs/d, Lbs/hr, Lbs/min, Lbs/sec, NLPH, NLPM, SLPM, SMPS, NMPS, SFPM, MMSCFD

Flow Rates for Insertion Flowmeter:

0 to 32,000 SFPM (163 NMPS) - Air at 70F & 1 ATM
To determine if an Insertion Flowmeter will operate properly, divide the maximum flow rate by the pipe area. The application is acceptable if the velocity is within the velocity range above.

Gas Pressure (maximum):
500 psig (34.5 barg)

Insertion Flowmeter:
Note: Pressure ratings stated for temperature of 100°F (38°C).

Temperature:
Std sensor: -40 to 250°F (-40 to 121°C)
HS Sensor: 32 to 400°F (0 to 204°C)
HT Sensor: 32 to 650°F (0 to 343°C)
Enclosure: -40 to 158°F (-40 to 70°C) without display, 32 to 140°F, (0 to 60°C) with display

Input Power:
24 VDC, ± 10%, 0.75 amp standard

Output:
Two isolated 4 to 20 mA outputs (1 for flow & 1 for temperature); isolated pulse output 0 to 100Hz, 10 volts p/p for flow (the pulse output can be used as an isolated solid state output for alarms)
RS232 for connecting a Palm Handheld or computer; RS422/ RS485-Modbus, Profibus-DP, DeviceNet or Ethernet

 


 

Physical Specs

Sensor material:

316 stainless steel standard; Hastelloy C276 optional Enclosure:
FM (U.S.) & FMc (CANADIAN) approved for Class I, II, III, Division 2, Groups A, B, C, D, E, F, G, T4A hazardous locations. NEMA 4X and CE approved.
Options: Remote NEMA 4X electronics enclosure with explosion-proof sensor j-box.

Cabling:
Remote enclosure: 5-conductor, 18 AWG, twisted, shielded, 100 feet maximum
Packing gland assembly: 125 psig (8.6 barg ) max. (See Figure 6)
High pressure (crank) retractor: NPT 600 psig (41.4 barg), ANSI 150 flange & ANSI 300 flange, no valve supplied. (See Figure 7)

 


 

Dimensional

Insertion Flowmeters:

Probe diameter: 1/2"

Equation for selecting Insertion Flowmeter probe length:

Probe length = ½ pipe ID (in inches) + 2" + thickness of insulation (if any) + dimension of retractor (if supplied). Round up to the next standard probe length available.